FDA Adverse Event Injury Summary report: N

HI-TORQUE BALANCE MIDDLEWEIGHT ELITE GUIDE WIRE

MDR report key: 2941063 · Received February 4, 2013

Report

Report Number
2024168-2013-00595
Event Type
Injury
Date Received
February 4, 2013
Date of Event
January 3, 2013
Report Date
January 10, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
DQX
PMA / PMN Number
K103101
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE SEPARATION WAS ABLE TO BE CONFIRMED. THE FAILURE TO ADVANCE COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. BASED ON A VISUAL AND DIMENSIONAL INSPECTION AND SCANNING ELECTRON MICROSCOPY (SEM) IMAGING OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE HAS BEEN RETURNED FOR INVESTIGATION. THE INVESTIGATION IS NOT COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE OF A PATIENT WITH STEMI OF THE DE NOVO RIGHT CORONARY ARTERY (RCA) USING A RIGHT RADIAL ACCESS, THE BALANCE MIDDLEWEIGHT (BMW) ELITE GUIDE WIRE COULD NOT ADVANCE THE ELBOW SHAPED VESSEL AND BECAME DETACHED. A .035 NON-ABBOTT GUIDE WIRE WAS USED TO STRAIGHTEN THE VESSEL; THE GUIDE WIRE PROXIMAL END WAS REMOVED FROM THE ANATOMY AND THE DISTAL FRAGMENT WAS REMOVED USING A SNARE. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA, HOWEVER, THE EVENT CAUSED A DELAY OF 15 MINUTES IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
47133 HI-TORQUE BALANCE MIDDLEWEIGHT ELITE GUIDE WIRE GUIDE WIRE DQX AV-TEMECULA-CT 2081301

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention GUIDE WIRE: 035 TERUMO