HI-TORQUE BALANCE MIDDLEWEIGHT ELITE GUIDE WIRE
Report
- Report Number
- 2024168-2013-00595
- Event Type
- Injury
- Date Received
- February 4, 2013
- Date of Event
- January 3, 2013
- Report Date
- January 10, 2013
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- DQX
- PMA / PMN Number
- K103101
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE SEPARATION WAS ABLE TO BE CONFIRMED. THE FAILURE TO ADVANCE COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. BASED ON A VISUAL AND DIMENSIONAL INSPECTION AND SCANNING ELECTRON MICROSCOPY (SEM) IMAGING OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE HAS BEEN RETURNED FOR INVESTIGATION. THE INVESTIGATION IS NOT COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
IT WAS REPORTED THAT DURING A PROCEDURE OF A PATIENT WITH STEMI OF THE DE NOVO RIGHT CORONARY ARTERY (RCA) USING A RIGHT RADIAL ACCESS, THE BALANCE MIDDLEWEIGHT (BMW) ELITE GUIDE WIRE COULD NOT ADVANCE THE ELBOW SHAPED VESSEL AND BECAME DETACHED. A .035 NON-ABBOTT GUIDE WIRE WAS USED TO STRAIGHTEN THE VESSEL; THE GUIDE WIRE PROXIMAL END WAS REMOVED FROM THE ANATOMY AND THE DISTAL FRAGMENT WAS REMOVED USING A SNARE. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA, HOWEVER, THE EVENT CAUSED A DELAY OF 15 MINUTES IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 47133 | HI-TORQUE BALANCE MIDDLEWEIGHT ELITE GUIDE WIRE | GUIDE WIRE | DQX | AV-TEMECULA-CT | 2081301 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | GUIDE WIRE: 035 TERUMO |