FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 2941048
·
Received January 31, 2013
Report
- Report Number
- 1720753-2013-01340
- Event Type
- Malfunction
- Date Received
- January 31, 2013
- Date of Event
- January 16, 2013
- Report Date
- January 31, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE FUSES AND CONNECTORS WERE RESEATED. THE INTERCONNECT CABLE AND THE SYSTEM INTERFACE BOARD WERE REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THE SYSTEM HAD AN INTERMITTENT BLACK IMAGE DURING CASE. THERE IS NO REPORT OF PT INJURY OR DEATH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 43475 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |