PFNA 10 130° L240 SST
Report
- Report Number
- 8030965-2013-00287
- Event Type
- Injury
- Date Received
- February 4, 2013
- Report Date
- November 9, 2012
- Manufacturer
- SYNTHES GMBH
- Product Code
- HSB
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. THE MEASURABLE DIMENSIONS WERE MEASURED AND WERE FOUND TO BE IN SPECIFICATION. WE ASSUME THAT DOMINANT DYNAMIC BENDING LOADS, DOUBLE SIDED. LED TO THE FATIGUE OF THE INVESTIGATED PFNA. OVER A PERIOD OF TIME THE IMPLANT HAD TO ABSORB AND NEUTRALIZE FORCES THAT MAY HAVE BEEN GREATER THAN THE TOLERABLE LOAD. POST OPERATIVE ACTIVITIES OF THE PATIENT DURING THE REHABILITATION IN COMBINATION WITH THE COMPLEX FRACTURE SITUATION AND AN INSTABLE FRACTURE SITUATION HAVE PLAYED A CERTAIN ROLE TOO. SEM OBSERVATIONS AND FINDINGS LED TO THE CONCLUSION THAT THE INVESTIGATED IMPLANT BROKE BECAUSE OF FATIGUE AND OVERLOAD.
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ON GOING; NO CONCLUSION COULD BE DRAWN. DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES, BUT IS SIMILAR TO DEVICE MARKETED IN THE USA.
DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: PATIENT SUFFERED A FALL AND WAS IMPLANTED WITH PFNA NAIL, BOLT, AND BLADE CONSTRUCT ON (B)(6) 2012. ON AN UNKNOWN POST-OPERATIVE DATE, PATIENT COMPLAINED OF HIP PAIN. EXAMINATION REVEALED FRACTURE IN THE PFNA. PATIENT RETURNED TO THE OR ON (B)(6) 2012 FOR EXPLANT OF HARDWARE. PATIENT WAS REVISED WITH LONGER PFNA NAIL CONSTRUCT. THIS IS 1 OF 3 REPORTS FOR THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 46221 | PFNA 10 130° L240 SST | PFNA NAIL | HSB | SYNTHES GMBH | 7782912 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Required Intervention |