FDA Adverse Event Injury Summary report: N

EASYSPRAY PRESSURE REGULATOR

MDR report key: 2940720 · Received June 16, 2006

Report

Report Number
2954761-2006-00017
Event Type
Injury
Date Received
June 16, 2006
Date of Event
May 23, 2006
Report Date
June 16, 2006
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FMF
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IN THIS SPECIFIC SURGERY FLOSEAL, TISSEEL AND COSEAL HAVE BEEN USED. TISSEEL AND COSEAL WERE APPLIED WITH A GAS-DRIVEN SPRAY DEVICE (EASYSPRAY). THE PNEUMOPERITONEUM IS THE RESULT OF THE SPRAY APPLICATION, AND NOT OF THE VARIETY OF PRODUCTS USED. BOTH THE INSTRUCTIONS FOR USE OF THE SPRAY DEVICE AND THE EASYSPRAY, ARE CAUTIONING AGAINST GAS-DRIVEN APPLICATION IN CLOSED CAVITIES. THIS CAUTION IS NOT INCORPORATED IN THE COSEAL SPRAY SYSTEM. CAUTION: ANY APPLICATION OF PRESSURIZED GAS MAY BE ASSOCIATED WITH A POTENTIAL RISK OF GAS EMPHYSEMA, TISSUE RUPTURE, GAS ENTRAPMENT WITH COMPRESSION, OR AIR EMBOLISM, WHICH MAY BE LIFE THREATENING. BE SURE TO TAKE APPROPRIATE MEASURES TO ADDRESS THESE RISKS BY OBSERVING THE RECOMMENDED MINIMUM SPRAYING DISTANCE (10 CM) AND MAXIMUM PRESSURE (2 BARS). CAUTION: SPRAYING INTO ENCLOSED BODY CAVITIES REQUIRES APPROPRIATE SAFETY MEASURES TO MAKE SURE THAT THE ABOVE MENTIONED RISKS WILL BE AVOIDED. THE PNEUMOPERITONEUM IS THE RESULT OF THE APPLICATION OF BOTH TISSEEL AND COSEAL WITH A SPRAY DEVICE, AND NOT THE RESULT OF THE PRESENCE OF THE PRODUCTS THEMSELVES. THERE IS NO CAUSAL RELATIONSHIP BETWEEN THE PNEUMOPERITONEUM AND THE USE OF FLOSEAL. A RETRAINING OF THE SURGEON IN ACCORDANCE WITH THE ABOVE-MENTIONED CAUTIONS IS REQUIRED. DISCUSSED NEED FOR ALLOWING GAS TO ESCAPE FROM ABDOMEN BEFORE CLOSING. PHYSICIAN AGREED WITH THIS INFORMATION.

Description of Event or Problem · 1

FLOSEAL APPLIED TO INFUNDIBULAR LIGAMENTS, THEN PERITONEUM PARTIALLY CLOSED W/PURSESTRING AND COSEAL SPRAYED INTO CUL DE SAC BEFORE TYING DOWN PURSESTRING. TISSEEL SPRAYED ON EDGE OF VAGINAL CUFF (PERITONEUM CLOSED AT THIS TIME). WHEN VAGINAL CUFF CLOSED, NOTED TISSEEL "BUBBLING." SURGEON WONDERED IF THIS WAS NORMAL, BUT FELT IT WAS GAS TRYING TO ESCAPE. AT 18 HRS, POST OP, PT. DISTENDED. FLAT PLATE OF ABDOMEN PERFORMED AND PNEUMOPERITONEUM NOTED, WHICH DR. (B)(6) FELT TO BE RELATED TO COSEAL, AND RETROPERITONEAL AIR ALSO NOTED, WHICH SHE FELT TO BE RELATED TO THE SPRAYING OF TISSEEL. THE FOLLOWING MORNING, UNDER CT GUIDED IMAGING, TWENTY 60CC SYRINGES OF AIR WAS REMOVED FROM ABDOMEN AND RETROPERITONEUM. DATE OF CT PROCEDURE (B)(6) 2006.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EASYSPRAY PRESSURE REGULATOR PRESSURE REGULATOR FMF BAXTER HEALTHCARE CORPORATION 0600012 NA

Patients

Seq Age Sex Outcome Treatment
1 53 YR Required Intervention