FDA Adverse Event Malfunction Summary report: N

DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS

MDR report key: 2940466 · Received February 4, 2013

Report

Report Number
3003787298-2013-00005
Event Type
Malfunction
Date Received
February 4, 2013
Report Date
January 9, 2013
Manufacturer
SYNTHES JENNERSVILLE
Product Code
HTJ
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ON GOING; NO CONCLUSION COULD BE DRAWN. A REVIEW OF THE DEVICE HISTORY RECORDS HAS BEEN REQUESTED.

Additional Manufacturer Narrative · 1

THE DEVICE HISTORY RECORD WAS REVIEWED AND (B)(4) FOUND ARE NOT RELEVANT TO THIS COMPLAINT CONDITION. NO OTHER ISSUES WERE FOUND DURING MANUFACTURE THAT ARE RELEVANT TO THIS COMPLAINT CONDITION. A MANUFACTURING EVALUATION WAS CONDUCTED. THE RETURNED PRODUCT IS COMPLETE AND IN GOOD OVERALL CONDITION. THE RETURNED PRODUCT WAS CHECKED DIMENSIONALLY AND FUNCTIONALLY AND PASSED. THERE ARE SMALL TOOL MARKS ON THE NEEDLE WHERE IT ATTACHES TO THE SLIDER. THESE MARKS ARE NOT ASSOCIATED WITH THE MANUFACTURE OF THIS PRODUCT. THIS EVALUATION FOUND NO LOOSE PARTS AND NOTHING THAT WOULD CAUSE AN INCORRECT MEASUREMENT.

Description of Event or Problem · 1

FACILITY STAFF RECEIVED A REPLACEMENT DEPTH GAUGE ABOUT A MONTH AGO ((B)(6) 2012) AND NOTICED UPON OPENING THE DEVICE THAT IT WAS NOT SITTING TOGETHER PROPERLY. WHEN THE STAFF TESTED THE DEPTH GAUGE THE SHAFT AND HANDLE WERE LOOSE. THE SHAFT SHOULD THREAD INTO THE HANDLE BUT THE SHAFT COULD NOT THREAD ALL THE WAY. THE DEPTH GAUGE WAS NOT MEASURING CORRECTLY, THE MEASUREMENTS WERE OFF ABOUT A MILLIMETER AND A HALF. THIS INCIDENT HAD NO PROCEDURE NOR PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
46952 DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS DEPTH GAUGE HTJ SYNTHES JENNERSVILLE 7102017

Patients

Seq Age Sex Outcome Treatment
1