FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 2940439 · Received February 3, 2013

Report

Report Number
1416980-2013-02730
Event Type
Malfunction
Date Received
February 3, 2013
Date of Event
December 22, 2012
Report Date
January 9, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS CONDITION IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF ANOTHER CONDITION. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS ONE OR MORE CYCLE ADVANCES TO THE NEXT FILL WHEN A SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2012. DURING NIGHT DRAIN CYCLE 2, THE PATIENT'S ULTRAFILTRATION READING WAS 1172ML. THIS INDICATES THAT THE HOME PATIENT (HP) DRAINED 1172ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1500ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.