FDA Adverse Event Malfunction Summary report: N

MOTOR EM210 LEGEND STYLUS TOUCH

MDR report key: 2940233 · Received February 2, 2013

Report

Report Number
1045254-2013-00067
Event Type
Malfunction
Date Received
February 2, 2013
Report Date
January 4, 2013
Manufacturer
MEDTRONIC INC.
Product Code
ERL
PMA / PMN Number
K081475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS NOT RETURNED. THE DEVICE WAS NOT RETURNED AND THEREFORE NO EVALUATION COULD BE PERFORMED. METHOD: NO TESTING METHODS PERFORMED.

Additional Manufacturer Narrative · 1

SERIAL #: (B)(4). DEVICE RETURNED: (B)(4) 2013. DEVICE EVALUATION: THE RETURNED DEVICE WAS EVALUATED. THE DRILL PASSED THE VISUAL INSPECTION, THE ATTACHMENT INSPECTION (ABILITY TO LOCK/UNLOCK), DIRECTION OF ROTATION INSPECTION, AND VOLTAGE CHECK. THE DRILL FAILED THE TEMPERATURE TEST AND THE SPEED OF ROTATION TEST. THE HIGHEST DETECTED TEMPERATURE DURING THE TEMPERATURE TEST WAS 113.2 DEG F, WITH A DETECTED 39.9 DEG RISE IN TEMPERATURE DURING THE TEST (DURATION = 5 MINS). DURING THE SPEED OF ROTATION TEST, WHEN THE DRILL WAS SET TO RUN AT 60K RPM, THE MEASURED SPEED OF THE DRILL WAS 56617 RPM, WHICH IS BELOW THE SPECIFIED ACCEPTABLE RANGE.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED TO MEDTRONIC THAT A LEGEND STYLUS TOUCH HANDPIECE WAS HEATING UP. THERE WAS NO REPORTED PATIENT OR USER IMPACT.