OT VERIO IQ METER
Report
- Report Number
- 3008382007-2013-02078
- Event Type
- Malfunction
- Date Received
- February 2, 2013
- Report Date
- January 9, 2013
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
FOLLOW UP #1 (03/26/2013)-DEVICE EVALUATION: THE METER AND TEST STRIPS INVOLVED WITH THIS COMPLAINT HAVE BEEN RETURNED TO LFS AND EVALUATED BY PRODUCT ANALYSIS (PA) ON 03/05/2013 AND (B)(4) 2013 RESPECTIVELY WITH THE FOLLOWING FINDINGS: THE METER AND TEST STRIPS PASSED ALL TESTING WITH NO FAULTS FOUND. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, ANOTHER FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING INACCURATE RESULT OF 29.4MMOL/L ON OTVIQ METER. RETESTED WITH FREESTYLE LITE - RESULT 5.6MMOL/L WITHIN 30 MINUTES. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED RESULTS DID NOT MEET LIFESCAN'S ACCURACY/PRECISION CRITERIA AND THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 45644 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3385256 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |