FDA Adverse Event Malfunction Summary report: N

ETAC SWIFT MOBIL

MDR report key: 2939583 · Received December 18, 2012

Report

Report Number
3008986539-2012-00076
Event Type
Malfunction
Date Received
December 18, 2012
Date of Event
September 1, 2012
Report Date
September 24, 2012
Manufacturer
ETAC SUPPLY CENTER AB
Product Code
INM
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THERE ARE TWO ORANGE BOLTS (ADJUSTMENT PINS) WHICH ATTACHES THE SEAT FRAME TO THE LOWER FRAME OF THE CHAIR. INVESTIGATION OF THE PRODUCT SHOWS THAT THE LEFT SIDE BOLT IS BROKEN INTO TWO PIECES. ETAC SUPPLY CENTER AB HAS TRIED TO RECREATE THE EVENT. IF THE BOLT IS FIXED IN PLACE IT'S NOT POSSIBLE TO RECREATE THE EVENT. BUT IF THE BOLT HAS NOT ENTERED PROPERLY INTO THE TWO FRAMES THE BOLT COULD BREAK. OUR CONCLUSION IS THAT HEIGHT ADJUST PIN HAS BEEN ASSEMBLED IN A WRONG WAY. THE MANUAL FOR SHOWER CHAIR SHOWS HOW TO ASSEMBLE THE BOLT CORRECTLY. RISK ANALYSIS HAS BEEN PERFORMED FOR THE PRODUCT. THE HAZARD WITH BOLTS BREAKING ARE HANDLE IN THE RISK ANALYSIS. THE RISK IS CONSIDERED ACCEPTABLE.

Description of Event or Problem · 1

A NURSE TOOK THE PT TO A SHOWER, THE NURSE WAS WASHING THE PT'S FEET (TOES). THERE WAS LOUD NOISE AND BOTH THE SHOWER CHAIR AND THE PT FELL TO THE FLOOR (LEFT SIDE). THE LEFT HAND SHOULDER AND THE SIDE WERE HURT, ALSO THE LEFT ARM STAYED UNDER THE HEMI ARM REST. THE SEAT HEIGHT WAS AT THE HIGHEST LEVEL. HEIGHT ADJUSTMENT PIN (ORANGE, LEFT SIDE) WAS BROKEN. X-RAY SHOWED NO INJURIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ETAC SWIFT MOBIL CHAIR, WITH CASTERS - INM INM ETAC SUPPLY CENTER AB 80209410 NA

Patients

Seq Age Sex Outcome Treatment
1 70 YR