FDA Adverse Event Malfunction Summary report: N

VENTED HUMIDIFICATION CHAMBER

MDR report key: 2938696 · Received January 31, 2013

Report

Report Number
9611451-2013-00070
Event Type
Malfunction
Date Received
January 31, 2013
Date of Event
January 15, 2013
Report Date
January 15, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE TWO COMPLAINT MR290 CHAMBERS WAS RECEIVED, ONE FROM LOT 120525 AND ONE FROM LOT 120530 (MANUFACTURED MAY 30, 2012). THE CHAMBERS WERE VISUALLY INSPECTED. RESULTS: IN BOTH CASES THE WATER BAG SPIKE WAS COMPLETELY SEPARATED FROM THE WATER FEEDSET TUBE. ADEQUATE GLUE WAS FOUND ON THE FEEDSET TUBE/SPIKE CONNECTION BUT THE GLUE WAS NOT BONDING. A LOT CHECK REVEALED ONE OTHER COMPLAINT OF THIS NATURE FOR LOT 120530, BUT NO OTHER COMPLAINTS FOR LOT 120525. CONCLUSION: WE WERE UNABLE TO DETERMINE THE CAUSE OF THE REPORTED FAULT. ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. THIS SUGGESTS THAT THE SPIKE BECAME LOOSE POST PRODUCTION, LIKELY AS A RESULT OF FAILURE OF THE GLUE BOND THAT JOINS THE SPIKE TO THE WATER FEEDSET TUBE. THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 CHAMBER STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARM." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT THE WATER FEED TUBE OF AN MR290 HUMIDIFICATION CHAMBER BECAME SEPARATED FROM THE BAG SPIKE DURING USE AND THAT THIS OCCURED IN TWO CASES. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
42800 VENTED HUMIDIFICATION CHAMBER AUTOFEED CHAMBER BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 120525

Patients

Seq Age Sex Outcome Treatment
1