DIMENSION® CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2013-00027
- Event Type
- Malfunction
- Date Received
- January 31, 2013
- Date of Event
- December 14, 2012
- Report Date
- January 10, 2013
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC
- Product Code
- MLM
- PMA / PMN Number
- K060502
- Removal / Correction Number
- 2517506-11-27-2012-018-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- SERVICE PERSONNEL
Narratives
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE BIASED LOW TACROLIMUS RESULTS IS UNKNOWN. THE ACCOUNT IS NOT USING THE FA3316 REAGENT LOT. THE REAGENT ISSUE IS UNDER INVESTIGATION BY SIEMENS HEALTHCARE DIAGNOSTICS INC. SIEMENS HEALTHCARE DIAGNOSTICS INC. ISSUED AN URGENT MEDICAL DEVICE RECALL ON 01-15-2013. THE RECALL LETTER CONFIRMED THE COMPLAINT OF LOW QC AND PATIENT SAMPLE RECOVERIES WITH THE DIMENSION TACR FLEX REAGENT CARTRIDGE LOT FA3316. THE RECALL LETTER INSTRUCTED CUSTOMERS TO IMMEDIATELY DISCARD ANY REMAINING REAGENT INVENTORY OF LOT FA3316. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
BIASED LOW RESULTS WERE OBTAINED ON TACROLIMUS (TACR) ON QC AND PATIENT SAMPLES IN A LOT CROSSOVER STUDY VERSUS AN ALTERNATE METHODOLOGY / ALTERNATE INSTRUMENT SYSTEM. THE PATIENT RESULTS WERE NOT REPORTED TO THE PHYSICIAN. THE BIAS WAS NOTED BY THE LABORATORY AND THEY DID NOT PUT THE LOT INTO USE. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED ON THE BASIS OF THE BIASED LOW TACR RESULTS. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE BIASED LOW TACROLIMUS RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 42864 | DIMENSION® CLINICAL CHEMISTRY SYSTEM | TACROLIMUS (TACR) FLEX® REAGENT CARTRIDGE | MLM | SIEMENS HEALTHCARE DIAGNOSTICS INC | FA3316 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |