COBE DIALYSIS MACHINE
MDR report key: 29365
·
Received December 12, 1995
Report
- Report Number
- 29365
- Event Type
- Injury
- Date Received
- December 12, 1995
- Date of Event
- November 30, 1995
- Report Date
- December 11, 1995
- Manufacturer
- CGH MEDICAL, INC.
- Product Code
- FII
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- OH, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
THE DIALYSIS MACHINE REMOVED 10 POUNDS OF FLUID IN TWO HRS RATHER THAN 9 POUNDS IN FOUR HOURS AS WAS PROGRAMMED. NO ALARM INDICATORS LIT AND AUTOTEST WAS NEGATIVE, ALSO WATER LEAKED. BIOMED DISCOVERED THE PRV3 DIAPHRAGM WAS TORN. THE PT FELT "FUNNY," WAS DIZZY AND PALE, SHE REQUIRED NORMAL SALINE SOLUTION INFUSION OF 200 CC.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COBE DIALYSIS MACHINE | DIALYSIS MACHINE | FII | CGH MEDICAL, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Required Intervention |