FDA Adverse Event Injury Summary report: N

COBE DIALYSIS MACHINE

MDR report key: 29365 · Received December 12, 1995

Report

Report Number
29365
Event Type
Injury
Date Received
December 12, 1995
Date of Event
November 30, 1995
Report Date
December 11, 1995
Manufacturer
CGH MEDICAL, INC.
Product Code
FII
Product Problem
Yes
Report Source
User Facility report
Reporter Location
OH, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

THE DIALYSIS MACHINE REMOVED 10 POUNDS OF FLUID IN TWO HRS RATHER THAN 9 POUNDS IN FOUR HOURS AS WAS PROGRAMMED. NO ALARM INDICATORS LIT AND AUTOTEST WAS NEGATIVE, ALSO WATER LEAKED. BIOMED DISCOVERED THE PRV3 DIAPHRAGM WAS TORN. THE PT FELT "FUNNY," WAS DIZZY AND PALE, SHE REQUIRED NORMAL SALINE SOLUTION INFUSION OF 200 CC.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COBE DIALYSIS MACHINE DIALYSIS MACHINE FII CGH MEDICAL, INC.

Patients

Seq Age Sex Outcome Treatment
1 75 YR Required Intervention