OT VERIO IQ METER
Report
- Report Number
- 3008382007-2013-01884
- Event Type
- Malfunction
- Date Received
- January 30, 2013
- Report Date
- January 7, 2013
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4): THE METER, USB CABLE, AND AC ADAPTER HAVE PASSED TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE REPRODUCED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
THE PATIENT'S PRODUCTS HAVE BEEN RETURNED TO LIFESCAN FOR EVALUATION; HOWEVER, THE EVALUATION PROCESS HAS NOT YET BEEN COMPLETED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2013 THE LAY USER/PATIENT IN (B)(6) CONTACTED LIFESCAN TO REPORT THE ONE TOUCH VERIO IQ METER WOULD NOT POWER ON. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS ISSUE. AS THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING THIS COMPLAINT IS BEING REPORTED. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 41281 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3314933 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |