HOMECHOICE
Report
- Report Number
- 1416980-2013-02391
- Event Type
- Malfunction
- Date Received
- January 30, 2013
- Date of Event
- November 15, 2012
- Report Date
- January 16, 2013
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS FALSE EMPTY DETECT AND USE ERROR WITH INITIAL DRAIN ALARM SETTING INAPPROPRIATELY PROGRAMMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.
DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2012 11:41:47. DURING NIGHT DRAIN CYCLE ONE, THE PATIENT'S ULTRAFILTRATION READING WAS 2735ML, INDICATING THE HOME PATIENT (HP) DRAINED 2735ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 250ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 40888 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |