CORE IMPACTION DRILL
MDR report key: 2935906
·
Received January 30, 2013
Report
- Report Number
- 0001811755-2013-00121
- Event Type
- Malfunction
- Date Received
- January 30, 2013
- Date of Event
- January 7, 2013
- Report Date
- January 15, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ND, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DURING THE EVALUATION OF THE DEVICE, CORRODED COMPONENTS WERE DISCOVERED WITHIN THE DEVICE, INCLUDING THE MOTOR CARTRIDGE, THE DRIVESHAFT ASSEMBLY AND THE BALANCED ROTOR. INCREASED FRICTION DUE TO CORROSION IS A PROBABLE CAUSE OF THE REPORTED OVERHEATING.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY, THE DRILL WAS OVERHEATING IN THE MIDDLE WHERE IT IS HELD BY THE USER. BACK-UP EQUIPMENT WAS USED TO COMPLETE THE PROCEDURE. NO DELAY, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 41116 | CORE IMPACTION DRILL | DRIVER, WIRE, AND BONE DRILL, MANUAL | DZJ | STRYKER INSTRUMENTS-KALAMAZOO |