FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 3000 SYSTEM
MDR report key: 2935892
·
Received January 2, 2013
Report
- Report Number
- 3008642652-2012-03423
- Event Type
- Malfunction
- Date Received
- January 2, 2013
- Date of Event
- December 5, 2012
- Report Date
- December 31, 2012
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. AS RECEIVED, THE TRUNK CABLE CONNECTOR HOUSING WAS BROKEN. THE CONNECTOR LOCKING NUT WAS CRACKED, WHICH WOULD PREVENT THE ELECTRODE BELT FROM PROPERLY SECURING INTO THE MONITOR. THE ROOT CAUSE FOR THE CRACKED LOCKING NUT CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY PHYSICAL ABUSE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED ELECTRODE BELT CONNECTOR.
Description of Event or Problem · 1
REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING, BELT SN (B)(4) WAS FOUND TO HAVE A DAMAGED TRUNK CABLE CONNECTOR. THE LAST PATIENT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1627 | LIFEVEST WCD 3000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |