FLOGARD
Report
- Report Number
- 1416980-2013-02318
- Event Type
- Malfunction
- Date Received
- January 30, 2013
- Date of Event
- November 1, 2012
- Report Date
- November 6, 2012
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4). INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. THIS DEVICE WAS EVALUATED ON-SITE BY A BAXTER FIELD SERVICE TECHNICIAN. UPON COMPLETION OF BAXTER'S INVESTIGATION, A FOLLOW-UP WILL BE SUBMITTED. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION.
(B)(4). EVALUATION SUMMARY: THE CUSTOMER REPORTED PROBLEM OF A FLOGARD INFUSION PUMP WITH AN "OCCLUSION CONSTANT ALARM" WAS CONFIRMED IN THE SAMPLE EVALUATION. THE CAUSE OF THE PROBLEM WAS A SAFETY CLAMP SHIM BEING OUT OF CALIBRATION. THE SAFETY CLAMP WAS RECALIBRATED ONSITE AT THE FACILITY BY A BAXTER FIELD SERVICE TECHNICIAN TO FIX THE PROBLEM.
ON (B)(6) 2012 THE CUSTOMER REPORTED A FLOGARD PUMP WITH AN OCCLUSION CONSTANT ALARM. IT IS UNKNOWN IF THIS EVENT OCCURRED DURING PATIENT USE. THERE WAS NO REPORT OF PATIENT INVOLVEMENT, PATIENT/USER INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 40374 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |