FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2935882 · Received January 30, 2013

Report

Report Number
1416980-2013-02318
Event Type
Malfunction
Date Received
January 30, 2013
Date of Event
November 1, 2012
Report Date
November 6, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. THIS DEVICE WAS EVALUATED ON-SITE BY A BAXTER FIELD SERVICE TECHNICIAN. UPON COMPLETION OF BAXTER'S INVESTIGATION, A FOLLOW-UP WILL BE SUBMITTED. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE CUSTOMER REPORTED PROBLEM OF A FLOGARD INFUSION PUMP WITH AN "OCCLUSION CONSTANT ALARM" WAS CONFIRMED IN THE SAMPLE EVALUATION. THE CAUSE OF THE PROBLEM WAS A SAFETY CLAMP SHIM BEING OUT OF CALIBRATION. THE SAFETY CLAMP WAS RECALIBRATED ONSITE AT THE FACILITY BY A BAXTER FIELD SERVICE TECHNICIAN TO FIX THE PROBLEM.

Description of Event or Problem · 1

ON (B)(6) 2012 THE CUSTOMER REPORTED A FLOGARD PUMP WITH AN OCCLUSION CONSTANT ALARM. IT IS UNKNOWN IF THIS EVENT OCCURRED DURING PATIENT USE. THERE WAS NO REPORT OF PATIENT INVOLVEMENT, PATIENT/USER INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
40374 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1