MDR report key: 2935852 · Received January 30, 2013

Report

Report Number
1416980-2013-02308
Event Type
Malfunction
Date Received
January 30, 2013
Date of Event
November 14, 2012
Report Date
November 14, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS REPAIRED ON-SITE AT THE CUSTOMER FACILITY BY A BAXTER FIELD SERVICE TECHNICIAN, AND THE CONDITION WAS CONFIRMED. THIS CONDITION IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF ANOTHER CONDITION. THE CAUSE WAS DETERMINED TO BE PHYSICAL DAMAGE TO THE POWER CORD STRAP. TO CORRECT THE CONDITION, THE POWER CORD STRAP WAS REPLACED.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE A DAMAGED POWER CORD STRAP. "THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY". THIS WAS NOT REPORTED BY THE CUSTOMER. "THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT." NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
41432 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1