FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2935809 · Received January 30, 2013

Report

Report Number
1416980-2013-02306
Event Type
Malfunction
Date Received
January 30, 2013
Date of Event
October 1, 2012
Report Date
October 1, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915523
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. UPON COMPLETION OF BAXTER'S INVESTIGATION, A FOLLOW-UP WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF FAILURE CODE 2 WAS CONFIRMED DURING DEVICE EVALUATION. THE ASSIGNABLE CAUSE OF THE FAILURE WAS IDENTIFIED AS A DAMAGED LATCH ROLLER. THE LATCH ROLLER WAS REPLACED TO CORRECT THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A FLOGARD INFUSION PUMP WITH A FAILURE CODE OF 2. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THERE WAS NO PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.