Additional Manufacturer Narrative · 1

INVESTIGATION FOUND THAT THE DIODE D7 WAS DETERMINED TO BE LEAKY AND WAS OUT OF THE SPECIFICATION. NEW PCB BOARD WAS REPLACED TO SOLVE THE ISSUE. (B)(4).

Description of Event or Problem · 1

INFO RECEIVED INDICATES THAT PUMP FAILS DIODE TEST WHICH IS THE CAUSE OF ERROR CODE 13 DURING TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1584 ENTERALITE INFINITY ENTERAL FEEDING PUMP LZH MOOG MED DEVICES GROUP INFINITY PUMP

Patients

Seq Age Sex Outcome Treatment
1