Report Number
1818910-2013-11675
Event Type
Injury
Date Received
January 30, 2013
Date of Event
January 23, 2013
Report Date
January 23, 2013
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
JDI
PMA / PMN Number
K980513
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PATIENT WAS REVISED TO ADDRESS HIP PAIN AND A MASS OF UNKNOWN TYPE IN GROIN. DOI 2004 - DOR (B)(6) 2013 (RIGHT HIP). THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATION SINCE ITS RELEASE FOR DISTRIBUTION. REVIEW OF THE DEVICE HISTORY RECORDS AND/OR A COMPLAINT DATABASE SEARCH WAS NOT POSSIBLE FOR THE FEMORAL HEAD AS THE PRODUCT AND LOT CODE REQUIRED WAS NOT PROVIDED. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT WITH THE INFORMATION AVAILABLE. BASED ON THE INABILITY TO IDENTIFY A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT OR ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS (B)(4) OTHER REPORTED INCIDENT AGAINST THE PROVIDED PRODUCT AND LOT COMBINATION. REVIEW OF THE DEVICE HISTORY RECORDS AND/OR A LOT SPECIFIC COMPLAINT DATABASE SEARCH WAS NOT POSSIBLE FOR THE UNKNOWN DEPUY FEMORAL HEAD AS THE PRODUCT AND LOT CODE REQUIRED WAS NOT PROVIDED. MEDICAL RECORDS WERE OBTAINED AND REVIEWED BY A MEDICAL PROFESSIONAL. FROM A MEDICAL PERSPECTIVE, BASED ON THE INFORMATION AVAILABLE, IT IS NOT POSSIBLE TO DETERMINE IF THE COMPLAINT IS PRODUCT RELATED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED.DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

PATIENT WAS REVISED TO ADDRESS HIP PAIN AND A MASS OF UNKNOWN TYPE IN GROIN.

Description of Event or Problem · 1

UPDATE REC'D 05/14/2014 - PFS AND MEDICAL RECORDS RECEIVED. A CORRECT DOI WAS GIVEN. AFTER REVIEW OF THE MEDICAL RECORDS THE REVISION OPERATIVE NOTE INDICATED METALLOSIS. THERE IS NO NEW ADDITIONAL INFORMATION THAT WOULD AFFECT THE EXISTING MDR DECISION.

Description of Event or Problem · 1

PPF ALLEGES METAL WEAR AND METALLOSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
40313 ARTICULEZE M HEAD 36MM +5 ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS JDI DEPUY ORTHOPAEDICS INC US 1116357

Patients

Seq Age Sex Outcome Treatment
1 75 YR Required Intervention