FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2935571 · Received January 30, 2013

Report

Report Number
1416980-2013-02294
Event Type
Malfunction
Date Received
January 30, 2013
Date of Event
January 9, 2013
Report Date
January 9, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915523
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF "F-38 ALARM" WAS CONFIRMED DURING DEVICE EVALUATION. THE ROOT CAUSE WAS DUE TO DEFECTIVE/INOPERATIVE FORCE SENSING RESISTORS. THE FORCE SENSING RESISTORS WERE REPLACED TO CORRECT THE REPORTED CONDITION.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER (B)(4) THAT A FLOGARD INFUSION PUMP HAD AN F-38 ALARM. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT; THEREFORE, NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
41558 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1