FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 2935418 · Received January 28, 2013

Report

Report Number
1720753-2013-01159
Event Type
Malfunction
Date Received
January 28, 2013
Date of Event
January 16, 2013
Report Date
January 28, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE POWER SUPPLY WAS ADJUSTED. THE SRAM AND THE SOFTWARE UPGRADE WERE INSTALLED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM WAS GETTING CHECK SUM ERROR AND WILL NOT BOOT UP. THERE IS NO REPORT OF PATIENT INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
37831 9600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9600

Patients

Seq Age Sex Outcome Treatment
1