FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2935321 · Received January 14, 2013

Report

Report Number
1720753-2013-00597
Event Type
Malfunction
Date Received
January 14, 2013
Date of Event
September 24, 2012
Report Date
January 14, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON-SITE INVESTIGATION. THE HIGH VOLTAGE TANK WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED MA SENSOR ERRORS DURING PT CASES. THERE IS NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
21234 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1