FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2935320 · Received January 14, 2013

Report

Report Number
1720753-2013-00599
Event Type
Malfunction
Date Received
January 14, 2013
Date of Event
December 31, 2012
Report Date
January 14, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON-SITE INVESTIGATION. THE FOLLOWING COMPONENTS WERE REPLACED: THE INFC BOARD, VIDEO CONTROLLER BOARD, IMAGE PROCESSOR BOARD, PS2 POWER SUPPLY AND LEFT MONITOR. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM BLACKED OUT AND SHUT DOWN. THERE IS NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
21203 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1