FDA Adverse Event Malfunction Summary report: N

8800

MDR report key: 2935252 · Received January 14, 2013

Report

Report Number
1720753-2013-00626
Event Type
Malfunction
Date Received
January 14, 2013
Date of Event
January 3, 2013
Report Date
January 14, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON-SITE INVESTIGATION. THE FLUORO FUNCTIONS BOARD, CINE DRIVE, AND THE SOFTWARE UPGRADE WERE INSTALLED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND OUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THE SYSTEM HAD A COMMUNICATION ERROR. THIS ERROR MAY RESULT IN A SYSTEM LOCK UP, NO BOOT, OR SHUT DOWN SITUATION DEPENDING ON WHEN THE LOSS OF COMMUNICATION OCCURRED. THERE IS NOT REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
21497 8800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 8800

Patients

Seq Age Sex Outcome Treatment
1