FDA Adverse Event Malfunction Summary report: N

2800

MDR report key: 2935246 · Received January 14, 2013

Report

Report Number
1720753-2013-00635
Event Type
Malfunction
Date Received
January 14, 2013
Date of Event
January 4, 2013
Report Date
January 14, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE SURGE SUPPRESSOR BOARD WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT POWER UP. THERE WAS NO PT INJURY OR DEATH REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
21330 2800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 2800

Patients

Seq Age Sex Outcome Treatment
1