FDA Adverse Event
Malfunction
Summary report: N
7700
MDR report key: 2935183
·
Received January 14, 2013
Report
- Report Number
- 9680959-2013-00119
- Event Type
- Malfunction
- Date Received
- January 14, 2013
- Date of Event
- December 17, 2012
- Report Date
- January 14, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS GMBH
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE FOOT SWITCH WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD PRODUCE FLUOROSCOPY X-RAY WITHOUT COMMAND. HOWEVER, THERE WAS NO PT INVOLVEMENT ASSOCIATED WITH THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 21302 | 7700 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS GMBH | 7700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |