FDA Adverse Event Malfunction Summary report: N

WINGED INFUSION SET, 20GA 3/4"

MDR report key: 2935053 · Received January 28, 2013

Report

Report Number
2245270-2013-00003
Event Type
Malfunction
Date Received
January 28, 2013
Date of Event
December 13, 2012
Report Date
January 28, 2013
Manufacturer
VYGON MANUFACTURING
Product Code
FMI
PMA / PMN Number
K873957
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

WHILE THE CUSTOMER DID NOT RETURN THE MFG DEVICES, THE CUSTOMER WAS ABLE TO RETURN FOUR UNOPENED DEVICES. VYGON PERFORMED A COMPLETE COMPLAINT INVESTIGATION USING THE UNOPENED SAMPLE AND A LOT HISTORY REVIEW. EXAMINATION OF THE UNOPENED DEVICES INDICATES THAT THE DEVICES FUNCTIONED AS INTENDED. VYGON PERFORMED A FUNCTIONAL TEST BY ATTACHING A WATER FILLED 60ML SYRINGE TO THE PROXIMAL END OF THE SETS AND PRESSURIZED THEM TO HELP DETERMINE THE ACTUAL LOCATION OF ANY LEAKAGE. NONE OF THE SETS LEAKED. WITHOUT HAVING THE ACTUAL SET TO EXAMINE AND AFTER CONDUCTING A THOROUGH INVESTIGATION INTO THIS CLAIM VYGON CANNOT CONFIRM A QUALITY PROBLEM WITH THIS LOT OF PRODUCT. ALL MFG STEPS WERE PERFORMED AND ALL IN-PROCESS QC TESTING PASS THE INSPECTION ACCEPTANCE QUALITY LEVEL (AOL). ACCORDING TO VYGON'S RECORDS, TWO HUNDRED PART WERE TESTED FOR PRESSURE AND OCCLUSION TESTING DURING MFG ALL WITH PASSING RESULTS. ALTHOUGH VYGON CANNOT CONFIRM A QUALITY PROBLEM WITH THIS PRODUCT, IT WILL CONTINUE TO BE VIGILANT FOR THIS TYPE OF COMPLAINT.

Description of Event or Problem · 1

NURSE PLACED HUBER NEEDLE IN PT'S PORT THEN ATTEMPTED TO FLUSH PORT WITH NORMAL SALINE. UPON PUSHING THE SALINE, IT WOULD SQUIRT BACKWARDS OUT OF TUBING AND ONTO THE PT. NURSE REMOVED THE SYRINGE FROM THE END OF TUBING AND UPON INSPECTION NOTICED THAT THE END OF TUBING WAS DEFECTIVE, EITHER CRACKED OR MIS-SHAPEN. THIS PRODUCT PROBLEM OCCURRED THREE MORE TIMES IN THE SAME WAY ON THE SAME DAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
36859 WINGED INFUSION SET, 20GA 3/4" HUBER INFUSION SET FMI VYGON MANUFACTURING MSY-2034 1111045

Patients

Seq Age Sex Outcome Treatment
1