WINGED INFUSION SET, 20GA 3/4"
Report
- Report Number
- 2245270-2013-00003
- Event Type
- Malfunction
- Date Received
- January 28, 2013
- Date of Event
- December 13, 2012
- Report Date
- January 28, 2013
- Manufacturer
- VYGON MANUFACTURING
- Product Code
- FMI
- PMA / PMN Number
- K873957
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NURSE
Narratives
WHILE THE CUSTOMER DID NOT RETURN THE MFG DEVICES, THE CUSTOMER WAS ABLE TO RETURN FOUR UNOPENED DEVICES. VYGON PERFORMED A COMPLETE COMPLAINT INVESTIGATION USING THE UNOPENED SAMPLE AND A LOT HISTORY REVIEW. EXAMINATION OF THE UNOPENED DEVICES INDICATES THAT THE DEVICES FUNCTIONED AS INTENDED. VYGON PERFORMED A FUNCTIONAL TEST BY ATTACHING A WATER FILLED 60ML SYRINGE TO THE PROXIMAL END OF THE SETS AND PRESSURIZED THEM TO HELP DETERMINE THE ACTUAL LOCATION OF ANY LEAKAGE. NONE OF THE SETS LEAKED. WITHOUT HAVING THE ACTUAL SET TO EXAMINE AND AFTER CONDUCTING A THOROUGH INVESTIGATION INTO THIS CLAIM VYGON CANNOT CONFIRM A QUALITY PROBLEM WITH THIS LOT OF PRODUCT. ALL MFG STEPS WERE PERFORMED AND ALL IN-PROCESS QC TESTING PASS THE INSPECTION ACCEPTANCE QUALITY LEVEL (AOL). ACCORDING TO VYGON'S RECORDS, TWO HUNDRED PART WERE TESTED FOR PRESSURE AND OCCLUSION TESTING DURING MFG ALL WITH PASSING RESULTS. ALTHOUGH VYGON CANNOT CONFIRM A QUALITY PROBLEM WITH THIS PRODUCT, IT WILL CONTINUE TO BE VIGILANT FOR THIS TYPE OF COMPLAINT.
NURSE PLACED HUBER NEEDLE IN PT'S PORT THEN ATTEMPTED TO FLUSH PORT WITH NORMAL SALINE. UPON PUSHING THE SALINE, IT WOULD SQUIRT BACKWARDS OUT OF TUBING AND ONTO THE PT. NURSE REMOVED THE SYRINGE FROM THE END OF TUBING AND UPON INSPECTION NOTICED THAT THE END OF TUBING WAS DEFECTIVE, EITHER CRACKED OR MIS-SHAPEN. THIS PRODUCT PROBLEM OCCURRED THREE MORE TIMES IN THE SAME WAY ON THE SAME DAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 36859 | WINGED INFUSION SET, 20GA 3/4" | HUBER INFUSION SET | FMI | VYGON MANUFACTURING | MSY-2034 | 1111045 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |