FDA Adverse Event Malfunction Summary report: N

NOVA MAX PLUS GLUCOSE MONITOR

MDR report key: 2935020 · Received January 25, 2013

Report

Report Number
3004193489-2013-00004
Event Type
Malfunction
Date Received
January 25, 2013
Date of Event
January 16, 2013
Report Date
January 24, 2013
Manufacturer
NOVA BIOMEDICAL CORP.
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

NOVA BIOMEDICAL AWAITS THE RETURN OF THE DEVICE FOR EVALUATION. SHOULD ANY SIGNIFICANT FINDINGS BE A RESULT OF THAT INVESTIGATION, A FOLLOW-UP REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED TO NOVA BIOMEDICAL THAT A CONSUMER RECEIVED A RESULT OF 523 MG/DL ON THEIR BLOOD GLUCOSE METER. THE CONSUMER IMMEDIATELY PERFORMED ANOTHER TEST USING THE SAME METER AND STRIPS GETTING THE FOLLOWING RESULT OF 187 MG/DL. DURING THE CALL TO CUSTOMER SUPPORT, THE CONSUMER STATED HAVING CONTROL TESTED THEIR TEST STRIPS WHEN THEY WERE OPENED AND THE TEST STRIPS CONTROL SOLUTION FELL INTO RANGE. ANOTHER CONTROL SOLUTION TEST WAS PERFORMED WHILE TROUBLESHOOTING SHOWING THE TEST STRIPS TO FALL IN RANGE. THE DIFFERENCE IN THE CONSUMER'S READINGS WAS DETERMINED TO BE CLINICALLY SIGNIFICANT. THE METER AND TEST STRIPS WILL BE RETURNED FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
36152 NOVA MAX PLUS GLUCOSE MONITOR GLUCOSE MONITOR NBW NOVA BIOMEDICAL CORP. NA 1020212207

Patients

Seq Age Sex Outcome Treatment
1 UNK