NOVAMAX LINK GLUCOSE MONITOR
Report
- Report Number
- 3004193489-2013-00002
- Event Type
- Malfunction
- Date Received
- January 25, 2013
- Date of Event
- January 3, 2013
- Report Date
- January 24, 2013
- Manufacturer
- NOVA BIOMEDICAL CORP
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- UNKNOWN
Narratives
NOVA BIOMEDICAL AWAITS THE RETURN OF THE DEVICE FOR EVALUATION. SHOULD ANY SIGNIFICANT FINDINGS BE A RESULT OF THAT INVESTIGATION, A FOLLOW-UP REPORT WILL BE FILED.
IT WAS REPORTED TO NOVA BIOMEDICAL THAT A CONSUMER RECEIVED A RESULT OF 160 MG/DL ON THEIR BLOOD GLUCOSE METER. THE CONSUMER IMMEDIATELY PERFORMED ANOTHER TEST USING THE SAME STRIPS GETTING AND ANOTHER NOVA METER GETTING A RESULT OF 40 MG/DL. DURING THE CALL TO CUSTOMER SUPPORT, THE CONSUMER STATED HAVING CONTROL TESTED THEIR TEST STRIPS WHEN THEY WERE OPENED AND THE TEST STRIPS CONTROL SOLUTION FELL INTO RANGE. ANOTHER CONTROL SOLUTION TEST WAS PERFORMED WHILE TROUBLESHOOTING SHOWING THE TEST STRIPS TO FALL IN RANGE. THE DIFFERENCE IN THE CONSUMER'S READINGS WAS DETERMINED TO BE CLINICALLY SIGNIFICANT. THE METER WILL BE RETURNED FOR EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 36237 | NOVAMAX LINK GLUCOSE MONITOR | GLUCOSE MONITOR | NBW | NOVA BIOMEDICAL CORP | NA | 1020512087 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |