PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE
Report
- Report Number
- 1419937-2013-00053
- Event Type
- Malfunction
- Date Received
- January 25, 2013
- Date of Event
- November 30, 2012
- Report Date
- November 30, 2012
- Manufacturer
- MEDELA, INC.
- Product Code
- HGX
- PMA / PMN Number
- K031614
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
A REPLACEMENT POWER SUPPLY WAS SENT TO THE CUSTOMER. IN FOLLOW-UP WITH THE CUSTOMER, THE CUSTOMER REPORTED THAT THERE WERE NO INJURIES AS A RESULT OF THE ISSUE. THE PRODUCT WAS RECEIVED ON (B)(4) 2012. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE RESULTING IN NEW, CHANGED OR CORRECTED INFORMATION, A FOLLOW-UP REPORT WILL BE FILED AT THAT TIME. A PRODUCT EVAL REVEALED THAT THE TRANSFORMER HOUSING WAS BROKEN, EXPOSING INNER ELECTRONICS, WHICH IS A SAFETY RISK. THIS TYPE OF FAILURE IS TYPICALLY ASSOCIATED WITH DROPPING THE UNIT ONTO A HARD SURFACE OR OTHER TYPE OF SEVERE IMPACT. THE ORIGINAL DESIGN TESTING INVOLVED DROPPING THE UNIT FROM A 1.0 M HEIGHT PER UL2601-1 2ND EDITION, PRODUCTION SAMPLES WERE DROPPED THREE TO FIVE TIMES FROM A HEIGHT OF 1.0 M AND THE HOUSINGS SHOWED DAMAGE AND CRACKING (ONLY AFTER AT LEAST THREE DROPS). THIS PRODUCT WAS RELEASED AS A RESULT OF CAPA (B)(4), WHICH WAS INITIATED FOR PUMP IN STYLE TRANSFORMER OVERHEATING/MELTING ISSUES FOR WHICH A FIELD ACTION SAFETY NOTIFICATION WAS INITIATED ON 04/04/2011. COMPLAINTS AGAINST THIS PRODUCT ARE CURRENTLY BEING INVESTIGATED ((B)(4)) IN ORDER TO DETERMINE ROOT CAUSE AND WILL CONTINUE TO BE MONITORED. A VISUAL EXAMINATION OF THE POWER SUPPLY REVEALED THE TRANSFORMER HOUSING WAS CRACKED OPEN, EXPOSING INNER ELECTRICAL CIRCUITRY. A VISUAL EXAMINATION OF THE CORD REVEALED NOTHING UNUSUAL. THE POWER SUPPLY WAS PLUGGED IN AND THE VOLTAGE ACROSS THE DC PLUG WAS MEASURED AT 14.07 VDC, WHICH IS NORMAL AND INDICATES THAT THE POWER SUPPLY IS PRODUCING THE CORRECT VOLTAGE. RESISTANCE ACROSS THE DC PLUG WAS MEASURED OPEN, WHICH INDICATES THAT THERE IS NOT A SHORT IN THE DC CORD. SMOKE, FIRE AND SPARKING WERE NOT OBSERVED WHILE THE POWER SUPPLY WAS PLUGGED IN. THE RESISTANCE ACROSS THE AC BLADES MEASURED AS 5.92 OHMS, WHICH IS NORMAL AND INDICATES THAT THE THERMAL FUSE DID NOT BLOW.
THE CUSTOMER REPORTED THAT THEIR PUMP IN STYLE DEVICE WOULD NOT POWER ON. UPON RECEIPT OF THE PRODUCT ON (B)(4) 2012, A BREACH IN THE TRANSFORMER HOUSING WAS NOTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 36101 | PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE | HGX | MEDELA, INC. | 57063 / 9207010 | REV L / 1511 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |