FDA Adverse Event Malfunction Summary report: N

TOTALCARE BED

MDR report key: 2934916 · Received December 21, 2012

Report

Report Number
1824206-2012-08489
Event Type
Malfunction
Date Received
December 21, 2012
Date of Event
December 5, 2012
Report Date
December 5, 2012
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND A SMASHED SIDERAIL CABLE AND CORROSION ON THE CIRCUIT BOARD DUE TO SMALL FRACTION OF FLUIDS ENTERING INTO THE SIDERAIL. THE SIDERAIL COVER DID NOT HAVE THE SEAL GASKET. THE TECHNICIAN REPLACED THE CABLE AND CIRCUIT BOARD TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATED THE LEFT USER CONTROL MODULE CONTROLS ARE OPERATING INTERMITTENTLY WHEN LOWERING THE SIDERAIL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE BED AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1