FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE BED
MDR report key: 2934916
·
Received December 21, 2012
Report
- Report Number
- 1824206-2012-08489
- Event Type
- Malfunction
- Date Received
- December 21, 2012
- Date of Event
- December 5, 2012
- Report Date
- December 5, 2012
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECHNICIAN FOUND A SMASHED SIDERAIL CABLE AND CORROSION ON THE CIRCUIT BOARD DUE TO SMALL FRACTION OF FLUIDS ENTERING INTO THE SIDERAIL. THE SIDERAIL COVER DID NOT HAVE THE SEAL GASKET. THE TECHNICIAN REPLACED THE CABLE AND CIRCUIT BOARD TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
INFORMATION RECEIVED INDICATED THE LEFT USER CONTROL MODULE CONTROLS ARE OPERATING INTERMITTENTLY WHEN LOWERING THE SIDERAIL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE BED | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |