ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-01346
- Event Type
- Malfunction
- Date Received
- January 30, 2013
- Report Date
- January 19, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION FOLLOW-UP #1 SUBMITTED 03/19/2013: THE DEVICE WAS RETURNED TO ANIMAS AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING RESULTS: TESTING CONFIRMED THE DOWN AND UP KEYS ARE INTERMITTENTLY RESPONDING TO USER INPUT AND REQUIRE EXCESSIVE FORCE TO ACTIVATE. CONTAMINATION WAS FOUND UNDER THE UP, DOWN AND CONTRAST KEY CONTACTS. UNRELATED TO THIS COMPLAINT, THE BLACK BOX SHOWS EVIDENCE OF TIME AND DATE RESETTING TO THE DEFAULT AFTER POWER ON RESETS ON 01/19/2013. REMOVAL OF THE PUMP COVER AND A VISIBLE INSPECTION CONFIRMED THE INTERNAL BATTERY WAS LEAKING.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER STATED THE DOWN ARROW KEYPAD BUTTON WAS INTERMITTENTLY UNRESPONSIVE AND REQUIRED MULTIPLE PRESSES TO ENGAGE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 41351 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 YR |