FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2934841 · Received January 4, 2013

Report

Report Number
1720753-2013-00158
Event Type
Malfunction
Date Received
January 4, 2013
Date of Event
December 19, 2012
Report Date
January 4, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE HARD DRIVE WAS REPLACED AND THE SYSTEM SOFTWARE WAS RELOADED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WILL INTERMITTENTLY NOT RECALL SAVED IMAGES. THERE IS NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
5779 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1