ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-01340
- Event Type
- Malfunction
- Date Received
- January 30, 2013
- Report Date
- January 18, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP # 1 DATE OF SUBMISSION 02/20/2013 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE PEELING AT THE OK BUTTON. A DAMAGED KEYPAD WILL ALLOW CONTAMINATION TO PERMEATE THE BUTTON CONTACTS NEGATIVELY IMPACTING THE BUTTON FUNCTION. THE DAMAGED KEYPAD IS OBVIOUS AND DETECTABLE AND SHOULD ALERT THE USER TO DISCONTINUE USE OF THE PUMP. DURING TESTING, THE UP ARROW AND DOWN ARROW KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY UNRESPONSIVE; THE OK AND CONTRAST KEYPAD BUTTONS RESPONDED APPROPRIATELY. THERE WAS EVIDENCE OF CONTAMINATION FOUND UNDER ALL KEY CONTACTS.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGE PRIOR TO DAMAGE) ISSUE. THE REPORTER STATED THE UP, DOWN, AND OK KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE AND REQUIRED MULTIPLE HARD PRESSES TO ENGAGE. THE REPORTER NOTED THE KEYPAD RUBBER WAS PEELING. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 40977 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |