FDA Adverse Event Malfunction Summary report: N

6800

MDR report key: 2934733 · Received January 4, 2013

Report

Report Number
1720753-2013-00208
Event Type
Malfunction
Date Received
January 4, 2013
Date of Event
December 19, 2012
Report Date
January 4, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND ADJUSTED THE 5 VOLT POWER SUPPLY. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
5765 6800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 6800

Patients

Seq Age Sex Outcome Treatment
1