FDA Adverse Event
Malfunction
Summary report: N
TRILOGY O2
MDR report key: 2934714
·
Received January 23, 2013
Report
- Report Number
- 2518422-2013-00109
- Event Type
- Malfunction
- Date Received
- January 23, 2013
- Date of Event
- January 8, 2013
- Report Date
- January 8, 2013
- Manufacturer
- RESPIRONICS INC.
- Product Code
- CBK
- PMA / PMN Number
- K093905
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DURING THE EVAL OF THE DEVICE AT THE MFR'S SERVICE CENTER, "SERVICE REQUIRED" CODES WERE FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SENSOR BOARD AND OXYGEN BLENDING MODULE BOARD WERE REPLACED TO ADDRESS THE ISSUE.
Description of Event or Problem · 1
A VENTILATOR WAS RETURNED TO THE MFR FOR SERVICE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 33274 | TRILOGY O2 | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | RESPIRONICS INC. | 1040004 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |