DIMENSION® VISTA CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2013-00036
- Event Type
- Malfunction
- Date Received
- January 30, 2013
- Date of Event
- January 5, 2013
- Report Date
- January 5, 2013
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- KXT
- PMA / PMN Number
- K112142
- Removal / Correction Number
- 2517506-03-06-2013-003-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY DEPRESSED DIGOXIN RESULTS IS UNKNOWN. THE CALIBRATOR LOT ASSOCIATED WITH THE DEPRESSED DIGOXIN RESULTS IS UNDER INVESTIGATION BY SIEMENS HEALTHCARE DIAGNOSTICS INC. THE ACCOUNT HAS NOT PLACED THE CALIBRATOR LOT INTO USE AND IS RUNNING DIGOXIN SAMPLES ON AN ALTERNATE DIMENSION VISTA(R) SYSTEM WITH DIGOXIN CALIBRATED WITH AN ALTERNATIVE LOT OF DRUG 4 CALIBRATOR. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS.
ORIGINAL MDR FILED: (B)(4) 2013. ADDITIONAL INFORMATION: SIEMENS HEALTHCARE DIAGNOSTICS INC. HAS CONFIRMED CUSTOMER COMPLAINTS OF LOW QC AND PATIENT RECOVERY FOR DIGOXIN WHEN THIS CALIBRATOR LOT IS USED TO CALIBRATE THE DIGXN METHOD. INTERNAL TESTING HAS CONFIRMED AN AVERAGE 15% LOW BIAS IN THE THERAPEUTIC RANGE OF 0.9 TO 2.0 NG/ML. THE CAUSE OF THE BIAS LOW IN DIGOXIN RECOVERIES IS CALIBRATOR INACCURACY. SIEMENS HEALTHCARE DIAGNOSTICS INC. CONDUCTED A VOLUNTARY RECALL FOR DIMENSION VISTA DRUG 4 CAL (KC460) LOT 2KD052 AND DRUG 4 CAL (KC460A) LOT 2KD053. AN URGENT MEDICAL DEVICE RECALL, COMMUNICATION #13-13, WAS ISSUED IN FEBRUARY 2013 TO IMPACTED CUSTOMERS. CUSTOMERS WERE INSTRUCTED TO IMMEDIATELY DISCONTINUE USE AND TO DISCARD ANY REMAINING INVENTORY OF DIMENSION VISTA DRUG 4 CAL (KC460) LOT 2KD052 AND DRUG 4 CAL (KC460A) LOT 2KD053.
BIASED LOW DIGOXIN RESULTS WERE OBTAINED ON QC AND PATIENT SAMPLES IN A LOT CROSS OVER STUDY. THE PATIENT RESULTS WERE NOT REPORTED TO THE PHYSICIAN. AFTER A TREND LOW WAS NOTED BY THE LABORATORY, THE SAMPLES WERE REPEATED ON THE ALTERNATE DIMENSION VISTA SYSTEM CALIBRATED WITH A DIFFERENT LOT OF CALIBRATOR AND HIGHER RESULTS WERE OBTAINED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED AS A RESULT OF THE FALSELY DEPRESSED DIGOXIN RESULTS. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY DEPRESSED DIGOXIN RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 40893 | DIMENSION® VISTA CLINICAL CHEMISTRY SYSTEM | DRUG 4 CALIBRATOR | KXT | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW | 2KD053 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |