FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 2934561
·
Received January 10, 2013
Report
- Report Number
- 1720753-2013-00476
- Event Type
- Malfunction
- Date Received
- January 10, 2013
- Date of Event
- December 28, 2012
- Report Date
- January 10, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE REP EVALUATED THE SYSTEM AND REPLACED THE SINGLE BOARD COMPUTER, DISPLAY ADAPTER BOARD, IMAGE PROCESSOR BOARD, THE SYSTEM INTERFACE BOARD, AND THE GENERATOR INTERFACE BOARD. THE SYSTEM OPERATES AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 16240 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |