HOMECHOICE
Report
- Report Number
- 1416980-2013-02248
- Event Type
- Malfunction
- Date Received
- January 30, 2013
- Date of Event
- December 23, 2012
- Report Date
- January 16, 2013
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
(B)(4). EVALUATION: THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MIN DRAIN VOLUME THRESHOLD. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.
DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2012, 23:10:59. DURING NIGHT DRAIN CYCLE SIX, THE PATIENT'S ULTRAFILTRATION READING WAS 1309ML, INDICATING THE HOME PATIENT (HP) DRAINED 1309ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1900ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 40859 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR |