FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2934531 · Received January 10, 2013

Report

Report Number
1720753-2013-00429
Event Type
Malfunction
Date Received
January 10, 2013
Date of Event
December 20, 2012
Report Date
January 10, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE HIGH VOLTAGE CABLE WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM INTERMITTENTLY WOULD NOT DISPLAY A LIVE IMAGE WHEN THE HIGH VOLTAGE CABLE WAS MOVED. THERE IS NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
16234 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1