EON MINI
Report
- Report Number
- 1627487-2013-12121
- Event Type
- Injury
- Date Received
- January 28, 2013
- Date of Event
- June 27, 2011
- Report Date
- January 3, 2013
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-002-R
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD ADVISORY AND IN A FIELD CORRECTION. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REFERENCE MFR REPORT #1627487-2013-12122. IT WAS REPORTED THE PATIENT HAS EXPERIENCED POCKET HEATING AT THE IPGS SITE SINCE THE IMPLANT DATE. PATIENT HAS BEEN SCHEDULED FOR SURGICAL INTERVENTION. ON 08/01/2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 38035 | EON MINI | SCS IPG | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3788 | 3376652 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR | Required Intervention | IMPLANT DATE:| SCS LEAD, MODEL 3219 |