840 VENTILATOR
MDR report key: 2934396
·
Received January 11, 2013
Report
- Report Number
- 8020893-2013-00082
- Event Type
- Malfunction
- Date Received
- January 11, 2013
- Date of Event
- January 3, 2013
- Report Date
- January 3, 2013
- Manufacturer
- COVIDIEN, NELLCOR PURITAN
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT AN 840 VENTILATOR HAD A BLANK SCREEN WHILE IN USE ON A PATIENT. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GRAPHIC USER INTERFACE CPU PCB. THE UNIT PASSED EXTENDED SELF-TESTING.