FDA Adverse Event Malfunction Summary report: N

SYSTEM 7 SAG SAW

MDR report key: 2934336 · Received January 30, 2013

Report

Report Number
0001811755-2013-00101
Event Type
Malfunction
Date Received
January 30, 2013
Date of Event
January 15, 2013
Report Date
January 16, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) 2013: THE REPORTED EVENT WAS CONFIRMED. DURING THE VISUAL INSPECTION, THE DIAGNOSTIC ENGINEER CONFIRMED A RED SOLID SUBSTANCE VISIBLE NEAR THE INDEXING BUTTON OF THE SAG HEAD. BASED ON THE FUNCTIONAL INSPECTION, THE SUBSTANCE IS A PRE-APPLIED LOCTITE THREAD LOCKER. THIS FAILURE IS MOST LIKELY ATTRIBUTED TO A MANUFACTURING ISSUE. THE LOANER DEVICE WAS REPAIRED AND PUT BACK INTO STRYKER STOCK.

Additional Manufacturer Narrative · 1

THE DEVICE IS NOT BEING RETURNED FOR EVALUATION AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE AND THE DEVICE IS RETURNED A FOLLOW UP REPORT WILL BE SUBMITTED. THE DEVICE IS NOT BEING RETURNED FOR EVALUATION AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE AND THE DEVICE IS RETURNED A FOLLOW UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM 7 SAGITTAL SAW WAS LEAKING AN UNKNOWN FLUID FOLLOWING CLEANING. THIS RESULTED IN A 1 HOUR DELAY TO A SURGICAL PROCEDURE. THERE WAS NO MEDICAL TREATMENT OR INTERVENTION REQUIRED AS A RESULT OF THIS EVENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM 7 SAGITTAL SAW WAS LEAKING AN UNKNOWN FLUID FOLLOWING CLEANING. THIS RESULTED IN A 1 HOUR DELAY TO A SURGICAL PROCEDURE. THERE WAS NO MEDICAL TREATMENT OR INTERVENTION REQUIRED AS A RESULT OF THIS EVENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
40410 SYSTEM 7 SAG SAW INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT HWE STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1