SYSTEM 7 SAG SAW
Report
- Report Number
- 0001811755-2013-00107
- Event Type
- Malfunction
- Date Received
- January 30, 2013
- Date of Event
- January 15, 2013
- Report Date
- January 16, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HWE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
THE REPORTED EVENT WAS CONFIRMED. DURING THE VISUAL INSPECTION, THE DIAGNOSTIC ENGINEER CONFIRMED A RED SOLID SUBSTANCE VISIBLE NEAR THE INDEXING BUTTON OF THE SAG HEAD. BASED ON THE FUNCTIONAL INSPECTION, THE SUBSTANCE IS A PRE-APPLIED LOCTITE THREADLOCKER. THE DEVICE WAS REPAIRED AND RETURNED TO STRYKER STOCK.
THE DEVICE IS NOT BEING RETURNED FOR EVALUATION AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE AND THE DEVICE IS RETURNED A FOLLOW UP REPORT WILL BE SUBMITTED. OTHER TEXT: THE DEVICE IS NOT BEING RETURNED FOR EVALUATION AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE AND THE DEVICE IS RETURNED A FOLLOW UP REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THE SYSTEM 7 SAGITTAL SAW WAS LEAKING AN UNKNOWN FLUID FOLLOWING CLEANING. THIS RESULTED IN A 1 HOUR DELAY TO A SURGICAL PROCEDURE. THERE WAS NO MEDICAL TREATMENT OR INTERVENTION REQUIRED AS A RESULT OF THIS EVENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.
IT WAS REPORTED THAT THE SYSTEM 7 SAGITTAL SAW WAS LEAKING AN UNKNOWN FLUID FOLLOWING CLEANING. THIS RESULTED IN A 1 HOUR DELAY TO A SURGICAL PROCEDURE. THERE WAS NO MEDICAL TREATMENT OR INTERVENTION REQUIRED AS A RESULT OF THIS EVENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 40300 | SYSTEM 7 SAG SAW | INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT | HWE | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |