2600

MDR report key: 2934293 · Received January 7, 2013

Report

Report Number
1720753-2013-00270
Event Type
Malfunction
Date Received
January 7, 2013
Date of Event
November 26, 2012
Report Date
January 7, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND RESEATED CIRCUIT BOARDS AND CONNECTORS, AND ADJUSTED THE 5 VOLT POWER SUPPLY. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM HAD MULTIPLE ERRORS AND SYMPTOMS AND INTERMITTENTLY LOCKED UP. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
8655 2600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 2600

Patients

Seq Age Sex Outcome Treatment
1