PFC SIGMARP CV TB/IN S3 12.5
Report
- Report Number
- 1818910-2013-11643
- Event Type
- Injury
- Date Received
- January 30, 2013
- Date of Event
- January 21, 2013
- Report Date
- January 21, 2013
- Manufacturer
- DEPUY IRELAND ? REG. # 9616671
- Product Code
- NJL
- PMA / PMN Number
- PP830055
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATION SINCE ITS RELEASE FOR DISTRIBUTION. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED POLYETHYLENE WEAR WITHOUT THE PRODUCT TO EXAMINE. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.
PATIENT WAS REVISED TO ADDRESS LACK OF MOTION. POLY WEAR WAS ALSO REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 41314 | PFC SIGMARP CV TB/IN S3 12.5 | TIBIAL KNEE INERT | NJL | DEPUY IRELAND ? REG. # 9616671 | 3219559 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Required Intervention |