FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 2934196 · Received January 3, 2013

Report

Report Number
1720753-2013-00122
Event Type
Malfunction
Date Received
January 3, 2013
Date of Event
December 14, 2012
Report Date
January 3, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE VIDEO BOARDS AND CONNECTORS WERE RESEATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE LEFT MONITOR WAS NOT WORKING. THIS ISSUE WILL CAUSE THE SYSTEM TO BE UNUSABLE DUE TO THE INABILITY TO SEE THE LIVE IMAGE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
4040 9600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9600

Patients

Seq Age Sex Outcome Treatment
1