FDA Adverse Event Malfunction Summary report: N

UNICEL® DXH 800 COULTER® CELLULAR ANALYSIS SYSTEM

MDR report key: 2933954 · Received January 29, 2013

Report

Report Number
1061932-2013-00121
Event Type
Malfunction
Date Received
January 29, 2013
Date of Event
December 20, 2012
Report Date
January 4, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K081930
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) PERFORMED COMPLETE BLOOD COUNT (CBC) FINE GAIN ADJUSTMENT AND FLOW RATE COMPENSATION PROCEDURES TO OPTIMIZE CBC APERTURE, AND ALSO PERFORMED SYSTEM OPERATION VERIFICATIONS PERTAINING TO CBC RECOVERY. THE CUSTOMER WAS ADVISED TO RECALIBRATE THE INSTRUMENT.

Description of Event or Problem · 1

THE CUSTOMER REPORTED ERRONEOUS HEMOGLOBIN AND HEMATOCRIT RESULTS, FOR ONE PATIENT, WITH CHECK FAILURES INVOLVING THE UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM, USED IN CONJUNCTION WITH S-CAL CALIBRATOR. THE RESULTS DID NOT CORRELATE ON TWO SAMPLE ANALYSES. REANALYZED RESULTS WERE CONSIDERED CORRECT AND RELEASED TO THE HOSPITAL. THE ERRONEOUS RESULTS WERE NOT RELEASED OUT OF THE LABORATORY. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT. BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
38899 UNICEL® DXH 800 COULTER® CELLULAR ANALYSIS SYSTEM COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1 36 YR